Category
Laboratory Water Purification Systems Ultra Pure Water System
Overview MSLUP03 Laboratory-GradeUltrapure Water Systems Highest Water Quality for Successful Analyses Product Feature P
Basic Info
Model NO. | MSLUP03 |
Product Description
MSLUP03 Laboratory-GradeUltrapure Water Systems
Highest Water Quality for Successful Analyses
- Applicable to various water quality, the case is with powder coating.
- The filter element adopts open mold injection technology and inside the machine
- Korean quick-insert filter column is easy to replace and maintain.
- The electrical appliance and waterways are completely separated to avoid circuit failure caused by moisture and water leakage.
- Online water quality monitoring system, immediately measure the output water quality.
- Automatic RO membrane anti-scaling flushing program, automatic stop when tap water break, automatic water breaking when machine stop, storage tank automatically fill water, automatic stopping when water is full.
- pipeline has function of upgrading water quality and quantity.
- Equipped with a portable TDS pen, you can test TDS content, conductance, water temperature anytime, anywhere.
- Double electric water intakes, two kinds water quality, ready to use. Pressure online monitoring.
- Open door at front and back, this make filters replacement easy&convenient, the replacement can be easily completed without using any tools.
- Double water intakes: RO water and UP water
MSLUP03 Ultrapure Water Machine | ||||
Type classification | Standard | Low Endotoxin | Low TOC | Comprehensive |
Production capacity | 10/20/30 | |||
The water requirements | Urban tap water, inlet water pressure: 0.15-0.5mpa, water temperature 5-40ºC | |||
RO water Parameter | ||||
Ion rejection rate | 97%-99% (when using a new RO membrane) | |||
Organic matter rejection rate | >99% when MW>200 Dalton | |||
Particle and bacterial rejection | > 99% | |||
UP water Parameter | ||||
Resistivity (25ºC) | 18.2 M Ω/cm @ 25 ºC | |||
Heavy metal ion | < 0.01 ppb | |||
Total organic carbon (TOC) * | <10 ppb | <3 ppb | ||
bacteria | < 0.01 cfu/ml | |||
Pyrogen (endotoxin) | N/A | < 0.001 Eu/ml | N/A | < 0.001 Eu/ml |
Particulate matter (> 0.1 microns) | <1/ml | |||
RNases | N/A | <1pg/ml | N/A | <1pg/ml |
DNases | N/A | <5pg/ml | N/A | <5pg/ml |
Dimensions and Weight | Length × width × height: 312×475×475mm;Weight: about 30Kg | |||
Electrical requirements | 100-240V, 50/60Hz, 48-72W | |||
The system configuration | Host machine (including 1 set of purification column)+12 l external pressure water storage tank +TDS pen + accessories package |
Application Field | |
Standard | All kinds of physical and chemical testing, biochemical analysis, blood examination, microanalysis, flushing special, HPLC, IC, GC, analysis experiment |
Low Endotoxin | Atomic absorption/emission spectrometry, mass spectrometry, TOC detection, environmental monitoring, biochip development, trace organic matter analysis |
Low TOC | High sensitivity ICP/MS, PCR, PPT level analysis, isotope analysis, molecular biology, life science, CDC, drug control institute, quality inspection institute, university scientific research and other standard laboratories and a variety of high-end precision instrument water |
Comprehensive | Molecular biology, Life sciences, gene research, cell culture, amino acid analysis, protein purification, toxicology, drug development, medical testing |